24
Fig. 13
Typical settings:
a) Set the Vertical switch to the CH 1 position
b) Press the HOLD button
c) Use the Menu button to select the AMAG OFF
sub-function and use the SELECT button to
switch it on ( AMAG ON.
d) Expand once as described in 9a or 9b.
e) The expanded signal waveform is displayed
roughly 3 div underneath the non-expanded
signal.
(12) Go-NoGo mode
The Go-NoGo sub-function is useful for the
assessment of signals. Each displayed "pixel" of
the signal waveform is "extended" upwards and
downwards by a specific amount. The following
"lengths" can be set: +/-0.4 div, +/-0.8 div or
+/-1.2 div. To select or set this sub-function,
proceed as follows:
a) In storage mode, press the MENU button until
"G-NG XH=>OH" appears at the top edge of
the screen in the middle.
b) Press the RECALL button once.
c) "0.4 div" appears instead of "XH=>OH". This
value can then be modified by pressing the
SELECT button (under MENU).
d) Then set the desired "assessment area".
e) Finally, to run the function, press the RECALL
button again.
(11) AMAG operating mode = ALT MAG
In this operating mode, both the "normal" stored signal waveform and the expanded signal waveform
of the relevant channel that was previously set, e.g. CH 1, are displayed simultaneously or alternately
= ALT.
To display CH 2, Vertical switch (18) must first be set to CH 2.
Note the following Table:
CH 1 CH 1-signal CH 1 signal + expanded CH 1 signal
CH 2 CH 2-signal CH 2 signal + expanded CH 2 signal
DUAL CH1- and CH2-Signal CH1 and CH2 signal, both channels
expanded
ADD CH1- + CH2-Signal The result of adding CH1 + CH2 at
the same time as the "expanded"
result.
Switch position Normal indication (not storage)