21
Fig. 9c Display of time base B
2.2.10 Storage mode
Storage mode functions as follows:
(1) "NORM" mode
a) Press the "STORAGE" button in the "STO-
RAGE MODE" panel.
b) "NORM" appears at the top edge of the
screen in the middle and the displayed signal
that is to be stored is then shown in the "real-
time" display.
c) All subsequent sweeps/displays are in slow
motion. If the Time/div switch is set to 0.1
µs/div, roughly 2 to 3 s elapses until a new
trace is painted and triggered.
d) If the Time/div switch is set to between 0.2
µs/div (in single-trace mode 5 µs/div) and 0.2
s/div (OS-3060 D: 0.1 µs/div to 0.2 s/div),
successive signal waveforms can be continu-
ously stored.
e) The "HOLD" button can be used to pause the
display, i.e. the signal is frozen in NORM
mode.
(2) EQUIV mode (equivalent)
If the Time/div switch for time base A is set to
between 0.2 µs/div and 2 µs/div (OS-3060: bet-
ween 0.1 µs/div and 2 µs/div), "EQUIV" (stands
for Equivalent) appears instead of NORM at the
top edge of the screen in the middle.
a) In this mode it is not possible to display the
left-hand end of the signal waveform (rising
or falling edge); the second cycle of the signal
must therefore be used for measurement pur-
poses.
b) Roughly 5 s elapse before a signal having a
frequency of less than 1 kHz is stored.
Fig. 10, EQUIV-MODE
(3) ROLL mode
The spot is slowly ”painted” from left to right.
This mode is especially suitable for signal wave-
forms having a frequency of less than roughly
100 Hz. Just press the "HOLD" button in order to
pause Roll mode.
ROLL MODE display
The signal display "rolls" from the right-hand
edge of the screen to the left. Sampling takes
place from right to left, i.e. signal changes are
first apparent at the right-hand edge of the
screen. Note the following figure.
90
10
90
90
10
DELAY TIME
Overbright "thicker" portion
of signal A trace
Fig. 9a, Display of time base A