2.2.4 Single-trace mode
Single-trace mode using the single time base
and internal triggering is a frequently used op-
erating mode. Use this mode if you only want to
observe one signal.
Because this oscilloscope is a dual-trace unit,
there are two particular ways of obtaining
single-trace operation.
2. Use the POSITION knob (16) or (17) to bring
the trace into the centre of the screen
3. Connect the signal lead carrying the signal to
be fed in to the channel you previously selec-
ted. Set the VOLTS/DIV switch so that the
signal amplitude takes up the entire height of
the screen.
4) To stop the signal moving, adjust the trigger
level knob (29) if necessary.
(b) Display for the deflection factor of A, B and X-Y, see also (1)(b)
Time / Div
> 0. 5 m s
„Blank“: X1, calibrated (not expanded) In X-Y mode, the usual*
: No interpolation ______ details dissappear ans “X-Y” appears instead
: Sinus (sin) interpolation X 10 expansion
: Linear (Lin) interpolation-
(c) Measurements using the cursor lines, see (1)(c)
a) If you want to read off the frequency on a
frequency counter at the same time as ob-
serving the signal, select channel 1 (CH 1) as
the signal feed. The signal applied to channel
1 can be picked off on a BNC socket at the rear
of the oscilloscope for further use, but at
reduced amplitude.
b) Channel 2 (CH 2) offers the facility to invert
the signal by means of the POSITION knob
(17), i.e. the signal is rotated through 180°.
Basic settings for single trace mode
1. Set up the oscilloscope for single trace mode as described below. Make sure that the trigger
SOURCE switch (27) and the channel that is used actually match (signal feed in channel 1 => switch
(27) set to CH 1)
Power switch (1) : set to ON
AC/GND/DC switch (10) or (11) : set tof AC
Vertical POSITION (16) or (17) : middle position and depressed
VARIABLE (14) or (15) : right-hand stop and depressed
Vertical MODE switch (18) : set to CH 1 or CH 2
Trigger MODE switch (26) : set to AUTO
Trigger SOURCE switch (24) : set to CH 1 or CH 2
Trigger LEVEL (29) : set to middle position
HOLDOFF knob (28) : set to NORM position = left-hand stop
Warning!
Do not exceed the maximum input values.
Never connect voltages in excess of 300 V DC or
212 V AC rms.